This paper presents the kinematic modeling, design, and experimental validation of MoviWE.Q, an innovative pseudo-omnidirectional electric-powered wheelchair (EPW) developed for indoor mobility. Conventional differential-drive EPWs struggle with maneuverability in confined spaces, while special-wheeled omnidirectional designs suffer from vibrations, noise, and low load capacity. To address these limitations, MoviWE.Q utilizes two swerve-drive locomotion units with standard conventional wheels. The platform integrates manual driving and caregiver-assisted driving via SensHB.Q, a cost-effective force-sensitive interface featuring three uniaxial load cells and compliant flexure joints. A complete kinematic model is formulated to manage actuation redundancy and determine system velocity twist limits. ISO 7176−5 tests and indoor navigation trials suggest excellent maneuverability in tight, cluttered environments. Furthermore, caregiver drive evaluations resulted in high usability scores, achieving a System Usability Scale (SUS) score of 83.1 and a low perceived workload (NASA-TLX score of 10.9).
Kinematic modeling, design, and testing of MoviWE.Q, a pseudo-omnidirectional electric-powered wheelchair for indoor mobility / Tagliavini, L., Botta, A., Colucci, G., Baglieri, L., Duretto, S., Toccaceli, L., Amodio, F., Quaglia, G.. - In: MECHANISM AND MACHINE THEORY. - ISSN 0094-114X. - 232:(2026), pp. 1-31. [10.1016/j.mechmachtheory.2026.106690]
Kinematic modeling, design, and testing of MoviWE.Q, a pseudo-omnidirectional electric-powered wheelchair for indoor mobility
Luigi Tagliavini;Andrea Botta;Giovanni Colucci;Lorenzo Baglieri;Simone Duretto;Lorenzo Toccaceli;Francesco Amodio;Giuseppe Quaglia
2026
Abstract
This paper presents the kinematic modeling, design, and experimental validation of MoviWE.Q, an innovative pseudo-omnidirectional electric-powered wheelchair (EPW) developed for indoor mobility. Conventional differential-drive EPWs struggle with maneuverability in confined spaces, while special-wheeled omnidirectional designs suffer from vibrations, noise, and low load capacity. To address these limitations, MoviWE.Q utilizes two swerve-drive locomotion units with standard conventional wheels. The platform integrates manual driving and caregiver-assisted driving via SensHB.Q, a cost-effective force-sensitive interface featuring three uniaxial load cells and compliant flexure joints. A complete kinematic model is formulated to manage actuation redundancy and determine system velocity twist limits. ISO 7176−5 tests and indoor navigation trials suggest excellent maneuverability in tight, cluttered environments. Furthermore, caregiver drive evaluations resulted in high usability scores, achieving a System Usability Scale (SUS) score of 83.1 and a low perceived workload (NASA-TLX score of 10.9).| File | Dimensione | Formato | |
|---|---|---|---|
|
261124 Kinematic modeling, design, and testing of MoviWE.Q Tagliavini Botta Colucci Baglieri Duretto Toccaceli Amodio Quaglia MMT_compressed.pdf
accesso aperto
Tipologia:
2a Post-print versione editoriale / Version of Record
Licenza:
Creative commons
Dimensione
14.21 MB
Formato
Adobe PDF
|
14.21 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11583/3018388
