Recent human-centred design studies suggest that acoustic noise could affect the physical use and psychological acceptance of a biomedical device. These aspects are especially relevant in the prosthetic field, in which device loudness is often related to rejection. The aim of the study is to inquire on the possibility to reduce the acoustic noise emitted by a robotic leg prosthesis by improving its casing. First, acoustic noise emissions are characterized experimentally using an anechoic chamber, both for the whole prosthesis, and for its actuator (i.e., noise source) in isolation. The characterizations show that the whole prosthesis including its casing amplify the actuator noise, and that noise emissions are concentrated within a certain frequency range. Based on these findings, the prosthesis casing has been redesigned to include a panel of Helmholtz resonator-based acoustic metamaterials as proof of concept, which attenuate respective noise emissions. Experimental validations show that the use of such metamaterials in the prosthesis casing can significantly reduce noise emissions without compromising on prosthesis size and weight.

Use of Metamaterials to Reduce Acoustic Noise Emissions from Lower Limb Prostheses: An Experimental Validation / Puliti, Marco; Driessen, Josephus; Vitale, Nicolo; Tessari, Federico; Traverso, Simone; Laffranchi, Matteo; De Michieli, Lorenzo; Shtrepi, Louena. - ELETTRONICO. - 2023:(2023), pp. 1-6. (Intervento presentato al convegno 2023 International Conference on Rehabilitation Robotics (ICORR) tenutosi a Singapore nel 24-28 September 2023) [10.1109/ICORR58425.2023.10304769].

Use of Metamaterials to Reduce Acoustic Noise Emissions from Lower Limb Prostheses: An Experimental Validation

Puliti, Marco;Tessari, Federico;Shtrepi, Louena
2023

Abstract

Recent human-centred design studies suggest that acoustic noise could affect the physical use and psychological acceptance of a biomedical device. These aspects are especially relevant in the prosthetic field, in which device loudness is often related to rejection. The aim of the study is to inquire on the possibility to reduce the acoustic noise emitted by a robotic leg prosthesis by improving its casing. First, acoustic noise emissions are characterized experimentally using an anechoic chamber, both for the whole prosthesis, and for its actuator (i.e., noise source) in isolation. The characterizations show that the whole prosthesis including its casing amplify the actuator noise, and that noise emissions are concentrated within a certain frequency range. Based on these findings, the prosthesis casing has been redesigned to include a panel of Helmholtz resonator-based acoustic metamaterials as proof of concept, which attenuate respective noise emissions. Experimental validations show that the use of such metamaterials in the prosthesis casing can significantly reduce noise emissions without compromising on prosthesis size and weight.
File in questo prodotto:
File Dimensione Formato  
Use_of_Metamaterials_to_Reduce_Acoustic_Noise_Emissions_from_Lower_Limb_Prostheses_An_Experimental_Validation.pdf

non disponibili

Tipologia: 2a Post-print versione editoriale / Version of Record
Licenza: Non Pubblico - Accesso privato/ristretto
Dimensione 795.97 kB
Formato Adobe PDF
795.97 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11583/2986466