In preparation for the upcoming deuterium-tritium campaign at the JET tokamak, the ex-vessel fission chamber neutron diagnostics and the neutron activation system will be calibrated in absolute terms at 14 MeV neutron energy, to a required accuracy of less than 10%. Two nominally identical DT neutron generators were chosen as the calibration sources, both of which were fully calibrated and characterized at the UK's National Physical Laboratory. The neutron activation method was adopted as a complementary method for the purpose of determining the absolute value of the neutron yield from the neutron generators and to provide a means of cross check for the active detection methods being employed. The work being presented here shows the derivation of the neutron emission rate from the neutron generators based upon experimental activation foil measurements.

Activation measurements in support of the 14 MeV neutron calibration of JET neutron monitors / Jednorog, S.; Laszynska, E.; Batistoni, P.; Bienkowska, B.; Cufar, A.; Ghani, Z.; Giacomelli, L.; Klix, A.; Loreti, S.; Mikszuta, K.; Packer, L.; Peacock, A.; Pillon, M.; Popovichev, S.; Rebai, M.; Rigamonti, D.; Roberts, N.; Tardocchi, M.; Thomas, D.; Subba, F.. - In: FUSION ENGINEERING AND DESIGN. - ISSN 0920-3796. - 125:(2017), pp. 50-56. [10.1016/j.fusengdes.2017.10.024]

Activation measurements in support of the 14 MeV neutron calibration of JET neutron monitors

Subba, F.
2017

Abstract

In preparation for the upcoming deuterium-tritium campaign at the JET tokamak, the ex-vessel fission chamber neutron diagnostics and the neutron activation system will be calibrated in absolute terms at 14 MeV neutron energy, to a required accuracy of less than 10%. Two nominally identical DT neutron generators were chosen as the calibration sources, both of which were fully calibrated and characterized at the UK's National Physical Laboratory. The neutron activation method was adopted as a complementary method for the purpose of determining the absolute value of the neutron yield from the neutron generators and to provide a means of cross check for the active detection methods being employed. The work being presented here shows the derivation of the neutron emission rate from the neutron generators based upon experimental activation foil measurements.
File in questo prodotto:
File Dimensione Formato  
53099_manuscript_revised.pdf

accesso aperto

Tipologia: 2. Post-print / Author's Accepted Manuscript
Licenza: Creative commons
Dimensione 2.39 MB
Formato Adobe PDF
2.39 MB Adobe PDF Visualizza/Apri
1-s2.0-S0920379617308918-main.pdf

non disponibili

Tipologia: 2a Post-print versione editoriale / Version of Record
Licenza: Non Pubblico - Accesso privato/ristretto
Dimensione 1.04 MB
Formato Adobe PDF
1.04 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11583/2986863