Instrumented Indentation Test (IIT) is a non-conventional mechanical characterization technique to evaluate hardness, Young modulus, creep and relaxation of materials. In the macro range, it represents a cheaper and faster alternative to conventional tensile-based tests. IIT is a metrological scale; thus, to establish traceability, calibration is essential. Frame compliance calibration is critical because it is a major contribution to the measurement uncertainty. This work discusses the limits of the current state-of-the-art and proposes a novel methodology for frame compliance calibration. The introduced approach demonstrates the source of common systematic errors in the mechanical characterization reported in the literature, i.e. edge effect, while first highlighting a relevant frame compliance nonlinearity. The proposed procedure is cost-effective and relies upon constitutive spring modelling of IIT and calibration of reference block by nanoindentation. Results show that the novel approach corrects systematic trends in the characterization and yields a relative measurement uncertainty of 5%.
Calibration of machine platform nonlinearity in Instrumented Indentation Test in the macro range / Kholkhujaev, J.; Maculotti, G.; Genta, G.; Galetto, M.. - In: PRECISION ENGINEERING. - ISSN 0141-6359. - ELETTRONICO. - 81:(2023), pp. 145-157. [10.1016/j.precisioneng.2023.02.005]
Calibration of machine platform nonlinearity in Instrumented Indentation Test in the macro range
Kholkhujaev J.;Maculotti G.;Genta G.;Galetto M.
2023
Abstract
Instrumented Indentation Test (IIT) is a non-conventional mechanical characterization technique to evaluate hardness, Young modulus, creep and relaxation of materials. In the macro range, it represents a cheaper and faster alternative to conventional tensile-based tests. IIT is a metrological scale; thus, to establish traceability, calibration is essential. Frame compliance calibration is critical because it is a major contribution to the measurement uncertainty. This work discusses the limits of the current state-of-the-art and proposes a novel methodology for frame compliance calibration. The introduced approach demonstrates the source of common systematic errors in the mechanical characterization reported in the literature, i.e. edge effect, while first highlighting a relevant frame compliance nonlinearity. The proposed procedure is cost-effective and relies upon constitutive spring modelling of IIT and calibration of reference block by nanoindentation. Results show that the novel approach corrects systematic trends in the characterization and yields a relative measurement uncertainty of 5%.File | Dimensione | Formato | |
---|---|---|---|
1-s2.0-S014163592300020X-main.pdf
accesso aperto
Tipologia:
2a Post-print versione editoriale / Version of Record
Licenza:
Creative commons
Dimensione
9.41 MB
Formato
Adobe PDF
|
9.41 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11583/2976427