Direct and online continuous monitoring of anaesthetic intake by the patient’s blood stream improves practices in the critical procedure of anaesthesia. Several systems have been proposed to address this challenge. Electrochemical techniques are highly appealing because of speed and accuracy, while blood spot sampling and fluidic devices are discouraged in this application, due to the necessity of having constant and continuous monitoring. To overcome this limitation, we developed a sensor suitable for measuring anaesthetics directly in patient’s vein. We present here, for the first time, a disposable and low-cost needle-shaped sensor based on a three-electrode electrochemical cell, made of pencil leads and platinum wire. The proposed sensor shows high linearity (99%) in measuring propofol directly in undiluted human serum at the temperature of 37°C in its therapeutic range (30 - 240 µM), with a limit of detection of 7.2 ± 3.0 µM.
Pencil graphite needle-shaped biosensor for anaesthetic monitoring in human serum / Aiassa, Simone; Yilmaz, Sinan; Carrara, Sandro; Demarchi, Danilo. - ELETTRONICO. - (2020), pp. 1-4. (Intervento presentato al convegno 2020 IEEE Sensors tenutosi a Rotterdam (NLD) nel October 25-28, 2020) [10.1109/SENSORS47125.2020.9278590].
Pencil graphite needle-shaped biosensor for anaesthetic monitoring in human serum
Simone Aiassa;Sandro Carrara;Danilo Demarchi
2020
Abstract
Direct and online continuous monitoring of anaesthetic intake by the patient’s blood stream improves practices in the critical procedure of anaesthesia. Several systems have been proposed to address this challenge. Electrochemical techniques are highly appealing because of speed and accuracy, while blood spot sampling and fluidic devices are discouraged in this application, due to the necessity of having constant and continuous monitoring. To overcome this limitation, we developed a sensor suitable for measuring anaesthetics directly in patient’s vein. We present here, for the first time, a disposable and low-cost needle-shaped sensor based on a three-electrode electrochemical cell, made of pencil leads and platinum wire. The proposed sensor shows high linearity (99%) in measuring propofol directly in undiluted human serum at the temperature of 37°C in its therapeutic range (30 - 240 µM), with a limit of detection of 7.2 ± 3.0 µM.File | Dimensione | Formato | |
---|---|---|---|
SENSORS2020(preprint).pdf
accesso aperto
Descrizione: Articolo principale post-print
Tipologia:
2. Post-print / Author's Accepted Manuscript
Licenza:
PUBBLICO - Tutti i diritti riservati
Dimensione
737.31 kB
Formato
Adobe PDF
|
737.31 kB | Adobe PDF | Visualizza/Apri |
SENSORS2020.pdf
non disponibili
Descrizione: Articolo principale post-print Versione Editoriale
Tipologia:
2a Post-print versione editoriale / Version of Record
Licenza:
Non Pubblico - Accesso privato/ristretto
Dimensione
1.31 MB
Formato
Adobe PDF
|
1.31 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11583/2846198