This paper describes a comparison calibration system applicable to meters for low electrolytic conductivity values. In this system, measurements are performed in a closed circuit with a flowing solution. This circuit contains a flow-through reference cell and a flow-through cell under calibration. Results of calibrations by substitution and by comparison at 3 mS/m (KCl aqueous solution) are compared. The result of a calibration by comparison at 0.3 mS/m is also reported.
Development of an in-line calibration system for flow-through cells for low electrolytic conductivity values / Elena, Orrù; Francesca, Durbiano; Ortolano, Massimo. - In: ACCREDITATION AND QUALITY ASSURANCE. - ISSN 0949-1775. - 19:(2014), pp. 11-16. [10.1007/s00769-013-1023-6]
Development of an in-line calibration system for flow-through cells for low electrolytic conductivity values
ORTOLANO, Massimo
2014
Abstract
This paper describes a comparison calibration system applicable to meters for low electrolytic conductivity values. In this system, measurements are performed in a closed circuit with a flowing solution. This circuit contains a flow-through reference cell and a flow-through cell under calibration. Results of calibrations by substitution and by comparison at 3 mS/m (KCl aqueous solution) are compared. The result of a calibration by comparison at 0.3 mS/m is also reported.Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11583/2532688
Attenzione
Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo