This paper describes design and test of a new impedance measurement system for non-linear devices that exhibits a seven decade range and works down to a frequency of 0.01 Hz. The system is specifically designed for electrochemical measurements so that it can work with a few millivolt stimuli still compensating electrochemical voltages of several volts, but the proposed architecture can be employed in many other fields where a flexible signal generation and analysis is required. The system employs an unconventional signal generator based on two PWM oscillators and an autocalibration system that allows uncertainties of less than 3% to be obtained over a range of 1 kΩ to 100 GΩ A synchronous demodulation processing allows the noise superposed to the low amplitude of the employed signals to be made negligible. The proposed system is able to work both as a stand-alone device or in connection with a personal computer. In this latter case, thanks to the instrument internal architecture, the personal computer can set almost any parameter, has access to the raw acquired data and can implement new data processing and analyses.
Hand Held Impedance Measurement System with Seven Decade Capability and Potentiostatic Function / Carullo, Alessio; Corbellini, Simone; Ferraris, Franco; Parvis, Marco; Vallan, Alberto; Angelini, EMMA PAOLA MARIA VIRGINIA; Grassini, Sabrina; Gallone, V.. - STAMPA. - 1:(2004), pp. 387-391. (Intervento presentato al convegno 21st IEEE Instrumentation and Measurement Technology Conference, IMTC/04 tenutosi a Como, Italy nel 18-20 May 2004) [10.1109/IMTC.2004.1351070].
Hand Held Impedance Measurement System with Seven Decade Capability and Potentiostatic Function
CARULLO, Alessio;CORBELLINI, SIMONE;FERRARIS, Franco;PARVIS, Marco;VALLAN, Alberto;ANGELINI, EMMA PAOLA MARIA VIRGINIA;GRASSINI, Sabrina;
2004
Abstract
This paper describes design and test of a new impedance measurement system for non-linear devices that exhibits a seven decade range and works down to a frequency of 0.01 Hz. The system is specifically designed for electrochemical measurements so that it can work with a few millivolt stimuli still compensating electrochemical voltages of several volts, but the proposed architecture can be employed in many other fields where a flexible signal generation and analysis is required. The system employs an unconventional signal generator based on two PWM oscillators and an autocalibration system that allows uncertainties of less than 3% to be obtained over a range of 1 kΩ to 100 GΩ A synchronous demodulation processing allows the noise superposed to the low amplitude of the employed signals to be made negligible. The proposed system is able to work both as a stand-alone device or in connection with a personal computer. In this latter case, thanks to the instrument internal architecture, the personal computer can set almost any parameter, has access to the raw acquired data and can implement new data processing and analyses.Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11583/1409729
Attenzione
Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo