Due to the increasing demand of climate change studies, traceability of measurements according to the International System of Units (SI) becomes fundamental to establish data quality and comparability in time. With the aim of guaranteeing the traceability, in October 2022, an on-site calibration of three thermometers operated by the Osservatorio meteorologico di Moncalieri was assessed. The calibration by comparison against traceable travelling standards was performed in a transportable thermal chamber manufactured by the Istituto Nazionale di Ricerca Metrologica (INRIM). Results showed that the main uncertainty contributions to the expanded uncertainty were the interpolation method, the chamber inhomogeneity and the repeatability of the sensors under calibration. This current calibration was compared with previous calibration in 2012 (on-site calibration) and in 2016 (laboratory calibration). The comparison analysis evidenced the drift effect on the thermometers and the importance of having an active calibration program to reduce this effect. Regarding the expanded uncertainty, both on-site calibrations presented the same order of magnitude and smaller than the correction values. This paper points out the advantage of performing an on-site calibration: since the calibration involves the same dataloggers and cables in the same environmental conditions, the calculated calibration curve represents more convincingly the real measurement conditions.

On-Site Calibration Procedure and Uncertainty Contributions on Air Temperature Sensors / Aranda, Natali Giselle; Merlone, Andrea. - In: INTERNATIONAL JOURNAL OF THERMOPHYSICS. - ISSN 0195-928X. - STAMPA. - 45:1(2024). [10.1007/s10765-023-03296-x]

On-Site Calibration Procedure and Uncertainty Contributions on Air Temperature Sensors

Aranda, Natali Giselle;Merlone, Andrea
2024

Abstract

Due to the increasing demand of climate change studies, traceability of measurements according to the International System of Units (SI) becomes fundamental to establish data quality and comparability in time. With the aim of guaranteeing the traceability, in October 2022, an on-site calibration of three thermometers operated by the Osservatorio meteorologico di Moncalieri was assessed. The calibration by comparison against traceable travelling standards was performed in a transportable thermal chamber manufactured by the Istituto Nazionale di Ricerca Metrologica (INRIM). Results showed that the main uncertainty contributions to the expanded uncertainty were the interpolation method, the chamber inhomogeneity and the repeatability of the sensors under calibration. This current calibration was compared with previous calibration in 2012 (on-site calibration) and in 2016 (laboratory calibration). The comparison analysis evidenced the drift effect on the thermometers and the importance of having an active calibration program to reduce this effect. Regarding the expanded uncertainty, both on-site calibrations presented the same order of magnitude and smaller than the correction values. This paper points out the advantage of performing an on-site calibration: since the calibration involves the same dataloggers and cables in the same environmental conditions, the calculated calibration curve represents more convincingly the real measurement conditions.
File in questo prodotto:
File Dimensione Formato  
s10765-023-03296-x.pdf

accesso aperto

Descrizione: On-Site Calibration Procedure and Uncertainty Contributions on Air Temperature Sensors
Tipologia: 2a Post-print versione editoriale / Version of Record
Licenza: Creative commons
Dimensione 1.01 MB
Formato Adobe PDF
1.01 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11583/2984797