Image-based personalized Finite Element Models (pFEM) could detect alterations in physiological deformation of human vertebral bodies, but their accuracy has been seldom reported. Meaningful validation experiments should allow vertebral endplate deformability and ensure well-controlled boundary conditions. This study aimed to (i) validate a new loading system to apply a homogeneous pressure on the vertebral endplate during vertebral body compression regardless of endplate deformation; (ii) perform a pilot study on human vertebral bodies measuring surface displacements and strains with Digital Image Correlation (DIC); (iii) determine the accuracy of pFEM of the vertebral bodies. Homogeneous pressure application was achieved by pressurizing a fluid silicone encased in a rubber silicone film acting on the cranial endplate. The loading system was validated by comparing DIC-measured longitudinal strains and lower-end contact pressures, measured on three homogeneous pseudovertebrae of constant transversal section at 2.0 kN, against theoretically calculated values. Longitudinal strains and contact pressures were rather homogeneous, and their mean values close to theoretical calculations (5% underestimation). DIC measurements of surface longitudinal and circumferential displacements and strains were obtained on three human vertebral bodies at 2.0 kN. Complete displacement and strain maps were achieved for anterolateral aspects with random errors ≤0.2 μm and ≤30 μstrain, respectively. Venous plexus and double curvatures limited the completeness and accuracy of DIC data in posterior aspects. pFEM of vertebral bodies, including cortical bone mapping, were built from computed tomography images. In anterolateral aspects, pFEM accuracy of the three vertebrae was: (i) comparable to literature in terms of longitudinal displacements (R2>0.8); (ii) extended to circumferential displacements (pooled data: R2>0.9) and longitudinal strains (zero median error, 95% error: <27%). Circumferential strains were overestimated (median error: 39%). The new methods presented may permit to study how physiological and pathologic conditions influence the ability of vertebral endplates/bodies to sustain loads.

Applying a homogeneous pressure distribution to the upper vertebral endplate: Validation of a new loading system, pilot application to human vertebral bodies, and finite element predictions of DIC measured displacements and strains / Baleani, M.; Fraterrigo, G.; Erani, P.; Rota, G.; Berni, M.; Taddei, F.; Schileo, E.. - In: JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS. - ISSN 1751-6161. - 140:(2023). [10.1016/j.jmbbm.2023.105706]

Applying a homogeneous pressure distribution to the upper vertebral endplate: Validation of a new loading system, pilot application to human vertebral bodies, and finite element predictions of DIC measured displacements and strains

Fraterrigo G.;
2023

Abstract

Image-based personalized Finite Element Models (pFEM) could detect alterations in physiological deformation of human vertebral bodies, but their accuracy has been seldom reported. Meaningful validation experiments should allow vertebral endplate deformability and ensure well-controlled boundary conditions. This study aimed to (i) validate a new loading system to apply a homogeneous pressure on the vertebral endplate during vertebral body compression regardless of endplate deformation; (ii) perform a pilot study on human vertebral bodies measuring surface displacements and strains with Digital Image Correlation (DIC); (iii) determine the accuracy of pFEM of the vertebral bodies. Homogeneous pressure application was achieved by pressurizing a fluid silicone encased in a rubber silicone film acting on the cranial endplate. The loading system was validated by comparing DIC-measured longitudinal strains and lower-end contact pressures, measured on three homogeneous pseudovertebrae of constant transversal section at 2.0 kN, against theoretically calculated values. Longitudinal strains and contact pressures were rather homogeneous, and their mean values close to theoretical calculations (5% underestimation). DIC measurements of surface longitudinal and circumferential displacements and strains were obtained on three human vertebral bodies at 2.0 kN. Complete displacement and strain maps were achieved for anterolateral aspects with random errors ≤0.2 μm and ≤30 μstrain, respectively. Venous plexus and double curvatures limited the completeness and accuracy of DIC data in posterior aspects. pFEM of vertebral bodies, including cortical bone mapping, were built from computed tomography images. In anterolateral aspects, pFEM accuracy of the three vertebrae was: (i) comparable to literature in terms of longitudinal displacements (R2>0.8); (ii) extended to circumferential displacements (pooled data: R2>0.9) and longitudinal strains (zero median error, 95% error: <27%). Circumferential strains were overestimated (median error: 39%). The new methods presented may permit to study how physiological and pathologic conditions influence the ability of vertebral endplates/bodies to sustain loads.
File in questo prodotto:
File Dimensione Formato  
JMBBM-S-22-01810-6-41.pdf

embargo fino al 02/02/2025

Tipologia: 1. Preprint / submitted version [pre- review]
Licenza: Creative commons
Dimensione 1.48 MB
Formato Adobe PDF
1.48 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11583/2982340