Aortic stiffness is an important clinical parameter for predicting cardiovascular events. Carotid–femoral pulse wave velocity (cf-PWV) has been proposed for performing this evaluation non-invasively; however, it requires dedicated equipment and experienced operators. We explored the possibility of measuring aortic stiffness using ultrasound scans of the abdominal aorta coupled with the Bramwell–Hill equation. Healthy subjects were investigated; measurements of cf-PWV were taken by arterial tonometry and aortic systo-diastolic pressure difference was estimated using a validated model. Pulsatility of an abdominal tract of aorta was assessed by automated processing of ultrasound scans. Through a Bland–Altmann analysis, we found large biases when estimating each parameter by applying the Bramwell–Hill equation to the measured values of the other two paramters (bias, ± 1.96 SD; PWV, about 2.1 ± 2.5 m/s; pulsatility, 12 ± 14%; pressure jump, 47 ± 55 mmHg). These results indicate that the two measures are not interchangeable, and that a large part of the bias is attributable to blood pressure estimation. Further studies are needed to identify the possible sources of bias between cf-PWV and aortic pulsatility.

Estimation of Aortic Stiffness with Bramwell–Hill Equation: A Comparative Analysis with Carotid–Femoral Pulse Wave Velocity / Mesin, Luca; Floris, Luca; Policastro, Piero; Albani, Stefano; Scacciatella, Paolo; Pugliese, Nicola; Masi, Stefano; Grillo, Andrea; Fabris, Bruno; Antonini-Canterin, Francesco. - In: BIOENGINEERING. - ISSN 2306-5354. - ELETTRONICO. - 9:7(2022), p. 265. [10.3390/bioengineering9070265]

Estimation of Aortic Stiffness with Bramwell–Hill Equation: A Comparative Analysis with Carotid–Femoral Pulse Wave Velocity

Luca Mesin;Piero Policastro;
2022

Abstract

Aortic stiffness is an important clinical parameter for predicting cardiovascular events. Carotid–femoral pulse wave velocity (cf-PWV) has been proposed for performing this evaluation non-invasively; however, it requires dedicated equipment and experienced operators. We explored the possibility of measuring aortic stiffness using ultrasound scans of the abdominal aorta coupled with the Bramwell–Hill equation. Healthy subjects were investigated; measurements of cf-PWV were taken by arterial tonometry and aortic systo-diastolic pressure difference was estimated using a validated model. Pulsatility of an abdominal tract of aorta was assessed by automated processing of ultrasound scans. Through a Bland–Altmann analysis, we found large biases when estimating each parameter by applying the Bramwell–Hill equation to the measured values of the other two paramters (bias, ± 1.96 SD; PWV, about 2.1 ± 2.5 m/s; pulsatility, 12 ± 14%; pressure jump, 47 ± 55 mmHg). These results indicate that the two measures are not interchangeable, and that a large part of the bias is attributable to blood pressure estimation. Further studies are needed to identify the possible sources of bias between cf-PWV and aortic pulsatility.
File in questo prodotto:
File Dimensione Formato  
AorticStiffness_BH.pdf

accesso aperto

Descrizione: Articolo principale - versione dell'editore
Tipologia: 2a Post-print versione editoriale / Version of Record
Licenza: Creative commons
Dimensione 1.03 MB
Formato Adobe PDF
1.03 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11583/2968558