The paper describes design, development and characterization of a measuring system for in-line temperature monitoring during freeze-drying of harmaceutical chemicals. T-type thermocouples with sub-micrometric thickness deposited via plasma-coated with SiOx film and embedded into the vial walls have been realised as prototypes and successfully tested. The measuring system does not alter the vial shape and its thermal conductivity and does not react with the lyophilized product thus avoiding any interference with the lyophilization process.
Smart thermocouples for in-line lyophilization monitoring / Iannucci, L.; Grassini, S.; Parvis, M.; Barresi, A. A.. - ELETTRONICO. - (2019), pp. 210-211. (Intervento presentato al convegno 7th European Drying Conference tenutosi a Torino, Italy nel 10-12 July 2019).
Smart thermocouples for in-line lyophilization monitoring
Iannucci L.;Grassini S.;Parvis M.;Barresi A. A.
2019
Abstract
The paper describes design, development and characterization of a measuring system for in-line temperature monitoring during freeze-drying of harmaceutical chemicals. T-type thermocouples with sub-micrometric thickness deposited via plasma-coated with SiOx film and embedded into the vial walls have been realised as prototypes and successfully tested. The measuring system does not alter the vial shape and its thermal conductivity and does not react with the lyophilized product thus avoiding any interference with the lyophilization process.File | Dimensione | Formato | |
---|---|---|---|
Eurodrying2019_Abstracts_210.pdf
accesso aperto
Descrizione: post print editoriale, con copertina
Tipologia:
2a Post-print versione editoriale / Version of Record
Licenza:
Creative commons
Dimensione
1.21 MB
Formato
Adobe PDF
|
1.21 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11583/2788072