Objective: The aim of this study is to mitigate intra-gradient coil eddy currents in a hybrid MRI-LINAC system. Methods: The tracks of the gradient coils are modified by cutting slits along the current flow direction. The electromagnetic model developed was first experimentally validated and then used to study the impacts of the slit conductors on the energized and surrounding coils. In this study, gradient coils were slit with different numbers of sub-Tracks and driven by a current with frequencies ranging from 100 Hz to 10 kHz. The proposed configuration was assessed by evaluating a number of system parameters, such as the gradient magnetic field, the power loss generated in the surrounding unenergized coil (hereafter referred to as passive coils), and the performance of the energized coil. Results: It was found that at a typical operating frequency of 1 kHz and compared with a conventional (no cut) split coil structure, the new coil pattern (with four slits) offered improved coil parameters. 1) The average power loss dissipated in the surrounding passive coil was significantly reduced by 85%, 2) the cuts largely reduced the secondary field generated by the eddy currents in the passive coil, which was reduced to about 4% of that produced by the uncut coil and, 3) the performance of the energized coil with slit tracks was significantly improved. Some typical gradient coil parameters, such as the figure of merit, efficiency (η), and \eta ^{2}/R (where η is the efficiency and R is the resistance), were improved by 8.0%, 11.9%, and 45.7%, respectively. Conclusion and Significance: The new slit coil structure is effective in mitigating intra-coil eddy current effects, which is an important issue in the MRI-LINAC system.

Mitigation of Intra-coil Eddy Currents in Split Gradient Coils in a Hybrid MRI-LINAC System / Tang, Fangfang; Freschi, Fabio; Repetto, Maurizio; Li, Yu; Liu, Feng; Crozier, Stuart. - In: IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. - ISSN 0018-9294. - 64:3(2017), pp. 725-732. [10.1109/TBME.2016.2573316]

Mitigation of Intra-coil Eddy Currents in Split Gradient Coils in a Hybrid MRI-LINAC System

FRESCHI, FABIO;REPETTO, MAURIZIO;
2017

Abstract

Objective: The aim of this study is to mitigate intra-gradient coil eddy currents in a hybrid MRI-LINAC system. Methods: The tracks of the gradient coils are modified by cutting slits along the current flow direction. The electromagnetic model developed was first experimentally validated and then used to study the impacts of the slit conductors on the energized and surrounding coils. In this study, gradient coils were slit with different numbers of sub-Tracks and driven by a current with frequencies ranging from 100 Hz to 10 kHz. The proposed configuration was assessed by evaluating a number of system parameters, such as the gradient magnetic field, the power loss generated in the surrounding unenergized coil (hereafter referred to as passive coils), and the performance of the energized coil. Results: It was found that at a typical operating frequency of 1 kHz and compared with a conventional (no cut) split coil structure, the new coil pattern (with four slits) offered improved coil parameters. 1) The average power loss dissipated in the surrounding passive coil was significantly reduced by 85%, 2) the cuts largely reduced the secondary field generated by the eddy currents in the passive coil, which was reduced to about 4% of that produced by the uncut coil and, 3) the performance of the energized coil with slit tracks was significantly improved. Some typical gradient coil parameters, such as the figure of merit, efficiency (η), and \eta ^{2}/R (where η is the efficiency and R is the resistance), were improved by 8.0%, 11.9%, and 45.7%, respectively. Conclusion and Significance: The new slit coil structure is effective in mitigating intra-coil eddy current effects, which is an important issue in the MRI-LINAC system.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11583/2679918
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo