For almost two decades, caesium fountain primary frequency standards (PFS) have represented the best realization of the definition of the second in the International System of units. Their accuracy has progressively improved with time, reaching a few parts in 10^16. In this paper, we present the accuracy evaluation of ITCsF2, the new Cs fountain PFS developed at the Italian National Metrological Institute, designed to be operated at cryogenic temperature to reduce the blackbody radiation shift. The short-term stability of the ITCsF2 fountain is 2 × 10−13τ−1/2 when operated at high atomic density, and the relative inaccuracy reaches 2.3 × 10−16. We also report four calibrations of International Atomic Time with a relative frequency agreement of (−1.7 ± 3.2) × 10−16, between ITCsF2 and the average of the other fountains operated in the world during the reference periods.

Accuracy evaluation of ITCsF2: a Nitrogen cooled Cesium Fountain / Filippo, Levi; Davide, Calonico; Claudio E., Calosso; Aldo, Godone; Salvatore, Micalizio; Costanzo, Giovanni Antonio. - In: METROLOGIA. - ISSN 0026-1394. - STAMPA. - 51:3(2014), pp. 270-284. [10.1088/0026-1394/51/3/270]

Accuracy evaluation of ITCsF2: a Nitrogen cooled Cesium Fountain

COSTANZO, Giovanni Antonio
2014

Abstract

For almost two decades, caesium fountain primary frequency standards (PFS) have represented the best realization of the definition of the second in the International System of units. Their accuracy has progressively improved with time, reaching a few parts in 10^16. In this paper, we present the accuracy evaluation of ITCsF2, the new Cs fountain PFS developed at the Italian National Metrological Institute, designed to be operated at cryogenic temperature to reduce the blackbody radiation shift. The short-term stability of the ITCsF2 fountain is 2 × 10−13τ−1/2 when operated at high atomic density, and the relative inaccuracy reaches 2.3 × 10−16. We also report four calibrations of International Atomic Time with a relative frequency agreement of (−1.7 ± 3.2) × 10−16, between ITCsF2 and the average of the other fountains operated in the world during the reference periods.
2014
File in questo prodotto:
File Dimensione Formato  
ITCsF2 Draft.pdf

accesso aperto

Tipologia: 1. Preprint / submitted version [pre- review]
Licenza: PUBBLICO - Tutti i diritti riservati
Dimensione 1.1 MB
Formato Adobe PDF
1.1 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11583/2528708
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo