This work is aimed at designing and constructing a loading rig for the synthetic hemi-pelvis; this system has been conceived with the goal of applying differently oriented articular forces in order to experimentally test the stress distribution and the stability of surgical reconstructions like, for example, hip arthroplasty or pelvic fixation. This device can be interfaced with a usual loading machine; it preserves the anatomy of the hemi-pelvis; it is simply constrained and it allows the simulation of all physiologic activities. Moreover, the visual accessibility of the peri-acetabular area has been guaranteed and this is imperative in order to be able to perform full-field analyses like a thermoelastic or photoelastic stress analysis. First experimental trials have shown a good repeatability of loading–unloading cycles (<1.2%), a low hysteresis (<2.4%) and a good dynamic behaviour (up to 10 Hz loading frequencies).

Human pelvis loading rig for static and dynamic stress analysis / Zanetti, Elisabetta; Bignardi, Cristina; Audenino, Alberto. - In: ACTA OF BIOENGINEERING AND BIOMECHANICS. - ISSN 1509-409X. - STAMPA. - 14:2(2012), pp. 61-66. [10.5277/abb120208]

Human pelvis loading rig for static and dynamic stress analysis

ZANETTI, ELISABETTA;BIGNARDI, CRISTINA;AUDENINO, Alberto
2012

Abstract

This work is aimed at designing and constructing a loading rig for the synthetic hemi-pelvis; this system has been conceived with the goal of applying differently oriented articular forces in order to experimentally test the stress distribution and the stability of surgical reconstructions like, for example, hip arthroplasty or pelvic fixation. This device can be interfaced with a usual loading machine; it preserves the anatomy of the hemi-pelvis; it is simply constrained and it allows the simulation of all physiologic activities. Moreover, the visual accessibility of the peri-acetabular area has been guaranteed and this is imperative in order to be able to perform full-field analyses like a thermoelastic or photoelastic stress analysis. First experimental trials have shown a good repeatability of loading–unloading cycles (<1.2%), a low hysteresis (<2.4%) and a good dynamic behaviour (up to 10 Hz loading frequencies).
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11583/2504231
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo