An innovative and modular system (LyoMonitor) for monitoring the primary drying of a lyophilization process in vials is illustrated: it integrates some commercial devices (pressure gauges, moisture sensor and mass spectrometer), an innovative balance and a manometric temperature measurement system based on an improved algorithm (DPE) to estimate sublimating interface temperature and position, product temperature profile, heat and mass transfer coefficients. A soft-sensor using a multipoint wireless thermometer can also estimate the previous parameters in a large number of vials. The performances of the previous devices for the determination of the end of the primary drying are compared. Finally, all these sensors can be used for control purposes and for the optimization of the process recipe; the use of DPE in a control loop will be shown as an example.

Monitoring of the primary drying of a lyophilization process in vials / Barresi, Antonello; Pisano, Roberto; Fissore, Davide; Rasetto, Valeria; Velardi, S. A.; Vallan, Alberto; Parvis, Marco; Galan, M.. - In: CHEMICAL ENGINEERING AND PROCESSING. - ISSN 0255-2701. - STAMPA. - 48:1(2009), pp. 408-423. [10.1016/j.cep.2008.05.004]

Monitoring of the primary drying of a lyophilization process in vials

BARRESI, Antonello;PISANO, ROBERTO;FISSORE, DAVIDE;RASETTO, VALERIA;VALLAN, Alberto;PARVIS, Marco;
2009

Abstract

An innovative and modular system (LyoMonitor) for monitoring the primary drying of a lyophilization process in vials is illustrated: it integrates some commercial devices (pressure gauges, moisture sensor and mass spectrometer), an innovative balance and a manometric temperature measurement system based on an improved algorithm (DPE) to estimate sublimating interface temperature and position, product temperature profile, heat and mass transfer coefficients. A soft-sensor using a multipoint wireless thermometer can also estimate the previous parameters in a large number of vials. The performances of the previous devices for the determination of the end of the primary drying are compared. Finally, all these sensors can be used for control purposes and for the optimization of the process recipe; the use of DPE in a control loop will be shown as an example.
File in questo prodotto:
File Dimensione Formato  
CEP_2009_Vol48(1b).pdf

accesso aperto

Tipologia: 2. Post-print / Author's Accepted Manuscript
Licenza: PUBBLICO - Tutti i diritti riservati
Dimensione 1.1 MB
Formato Adobe PDF
1.1 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11583/1861352
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo