This study presents a novel application of a dual-branch neural network to non-invasively estimate Systolic and Diastolic Blood Pressure (SBP and DBP) using fingertip electrocardiogram (ECG) and photoplethysmogram (PPG) signals, without the need for prior calibration. Addressing the gap in available databases, which often lack clear, unfiltered recordings suitable for non-clinical use, this research utilizes a newly preliminary developed single-subject dataset to enhance reliability and interpretability. Utilizing data collected from a single subject over two months, with 121 recordings of 30 s of synchronous ECG and PPG and relative blood pressure, the model incorporates both convolutional and LSTM branches to process and integrate spatial and temporal signal features effectively. The research demonstrates the neural network’s potential through a 10-fold cross-validation approach, achieving, in average, a mean absolute error of 3.82 ± 0.76 for SBP and 3.26 ± 0.76 for DBP across all folds. These results underscore the accuracy of the approach and highlight the need for further validation across a broader dataset to improve generalizability and reliability in non-clinical settings.
Dual-Branch Fusion Neural Network Model: A Preliminary Demonstration of Absolute, Calibration-Free Blood Pressure Estimation from ECG and PPG Signals / Delrio, Federico; Randazzo, Vincenzo; Cirrincione, Giansalvo; Pasero, Eros (SMART INNOVATION, SYSTEMS AND TECHNOLOGIES). - In: Neural Networks: Overview of Current Theories and ApplicationsSTAMPA. - [s.l] : Springer Nature, 2026. - ISBN 9789819540716. - pp. 41-50 [10.1007/978-981-95-4072-3_4]
Dual-Branch Fusion Neural Network Model: A Preliminary Demonstration of Absolute, Calibration-Free Blood Pressure Estimation from ECG and PPG Signals
Delrio, Federico;Randazzo, Vincenzo;Cirrincione, Giansalvo;Pasero, Eros
2026
Abstract
This study presents a novel application of a dual-branch neural network to non-invasively estimate Systolic and Diastolic Blood Pressure (SBP and DBP) using fingertip electrocardiogram (ECG) and photoplethysmogram (PPG) signals, without the need for prior calibration. Addressing the gap in available databases, which often lack clear, unfiltered recordings suitable for non-clinical use, this research utilizes a newly preliminary developed single-subject dataset to enhance reliability and interpretability. Utilizing data collected from a single subject over two months, with 121 recordings of 30 s of synchronous ECG and PPG and relative blood pressure, the model incorporates both convolutional and LSTM branches to process and integrate spatial and temporal signal features effectively. The research demonstrates the neural network’s potential through a 10-fold cross-validation approach, achieving, in average, a mean absolute error of 3.82 ± 0.76 for SBP and 3.26 ± 0.76 for DBP across all folds. These results underscore the accuracy of the approach and highlight the need for further validation across a broader dataset to improve generalizability and reliability in non-clinical settings.| File | Dimensione | Formato | |
|---|---|---|---|
|
2026_bookExtracted_WIRN2024_ABP estimation_Delrio.pdf
accesso riservato
Tipologia:
2a Post-print versione editoriale / Version of Record
Licenza:
Non Pubblico - Accesso privato/ristretto
Dimensione
451.56 kB
Formato
Adobe PDF
|
451.56 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
|
WIRN_2024___Absolute_Blood_Pressure_Prediction_With_Wide_Space_Time_Neural_Network.pdf
embargo fino al 05/04/2027
Tipologia:
2. Post-print / Author's Accepted Manuscript
Licenza:
Pubblico - Tutti i diritti riservati
Dimensione
775.07 kB
Formato
Adobe PDF
|
775.07 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11583/3010570