Using matrix algebra, a mathematical model is formulated for a particular type of external fixator with wires (system developed by Ilizarov) for the treatment of bone fractures. The mathematical model is used to give a linear estimate of the stiffness under lateral and axial loads in a representative number of practical conditions. Relative displacements of the bone ends at the fracture site are calculated not only in the common case of a gap, but also for various angles of inclined sliding contact; in this case, a realistic load is applied and nonlinear stiffening of the wires under transversal loads is iteratively taken into account.

Mechanical performance of external fixators with wires for the treatment of bone fractures. Part I: Load-displacement behavior / Delprete, Cristiana; Gola, Muzio. - In: JOURNAL OF BIOMECHANICAL ENGINEERING. - ISSN 0148-0731. - 115:1(1993), pp. 29-36.

Mechanical performance of external fixators with wires for the treatment of bone fractures. Part I: Load-displacement behavior

DELPRETE, CRISTIANA;GOLA, Muzio
1993

Abstract

Using matrix algebra, a mathematical model is formulated for a particular type of external fixator with wires (system developed by Ilizarov) for the treatment of bone fractures. The mathematical model is used to give a linear estimate of the stiffness under lateral and axial loads in a representative number of practical conditions. Relative displacements of the bone ends at the fracture site are calculated not only in the common case of a gap, but also for various angles of inclined sliding contact; in this case, a realistic load is applied and nonlinear stiffening of the wires under transversal loads is iteratively taken into account.
1993
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11583/1663604
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo